Device Recall TSX101A: AQUILION 16; System, XRay, Tomography, Computed 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Toshiba American Medical Systems Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    59690
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-3198-2011
  • 사례 시작날짜
    2011-06-22
  • 사례 출판 날짜
    2011-09-13
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2014-08-11
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System, x-ray, tomography, computed - Product Code JAK
  • 원인
    Toshiba america medical systems is initiating a field correction on their tsx-301b: aquilion premium; system, x-ray, tomography, computed, tsx-301a: aquilion one; system, x-ray, tomography, computed, tsx-20ia: aqulllon lb; system, x-ray, tomography, computed, and tsx-1o1a: aquilion 16; system, x-ray, tomography, computed due to several software issues that cause the scanners to function incorr.
  • 조치
    Toshiba America Medical Systems (TAMS) sent an URGENT MEDICAL DEVICE CORRECTION letter dated June 22, 2011, to all affected customers. Customers were instructed that a revised software to prevent occurrence of the problem would be installed on their system. Their Toshiba Service Representative would contact them for an appointment to install the software to their system when it became available. Customers were requested to take temporary measures until corrective measures were taken. 1) In dual scanoscopy, do not press the short key between the completion of the first scanoscopy and the start of the second scanoscopy. 2) In dual scanoscopy, do not click the (Scano Skip) button immediately before the completion of the first scanoscopy. 3) When the SplitCase function is set to ON, do not select two or more orders simultaneously from the appointment list. 4) When an image comment is to be preset by making a selection from the preset memo, do not select any image comment with more than 10 characters. Customers were asked to share the information with all users and reviewing radiologists as well as clinical engineering or Biomedical group at the facility. Customers were asked to complete and return the attached form and fax it to 877-349-3054. For any questions call (800) 421-1968 or contact your local Toshiba Representative at (800) 521-1968.

Device

  • 모델명 / 제조번호(시리얼번호)
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution
  • 제품 설명
    TSX-101A: AQUILION 16; System, X-Ray, Tomography, Computed || X-Ray Tomography
  • Manufacturer

Manufacturer

  • 제조사 주소
    Toshiba American Medical Systems Inc, 2441 Michelle Dr, P.O. Box 2068, Tustin CA 92781-2068
  • 제조사 모회사 (2017)
  • Source
    USFDA