Device Recall UniCel Dxl Access Immunoassay System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Beckman Coulter Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    36225
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1518-06
  • 사례 시작날짜
    2006-07-28
  • 사례 출판 날짜
    2006-09-23
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-03-29
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Immunoassay system - Product Code JJE
  • 원인
    The neck of the wash buffer cube can be extended higher than intended possibly causing the system to aspirate air instead of buffer before the buffer container is empty without the system immediately indicating that the supply is empty. this may lead to incorrect assay results.
  • 조치
    A Product Corrective Action (PCA) letter was sent on July 28, 2006 to all Dxl customers to inform them that due to a recent change in packaging it is possible that the neck of the wash buffer cube can be extended higher than intended. This condition can cause the system to aspirate air instead of buffer before the buffer container is empty without the system immediately indicating that the supply is empty. If they are running System software version 2.2.2 (distributed beginning March 2006) and greater it will detect the premature out-of-buffer state, thus preventing the potential for erroneous results. The letter also informed the customers that if they run system software version 2.2.1 or earlier with these wash buffer lot numbers, the premature out-of-buffer state will not be detected immediately, leading to the potential for erroneous results. They were informed that Installation of software version 2.2.2 or greater is mandatory. To ensure immediate upgrade to version 2.2.2, customers are to contact the Clinical Support Center or their local Beckman Coulter representative. After system software version 2.2.2 or greater is installed, the system will detect the out-of-buffer state, stop sample processing, and suppress results. One of the following Event Log messages will be posted: ''Air or excessive restriction has compromised wash buffer delivery'' or ''Air detected in the wash buffer supply line'' To prevent out-of-buffer events users are to complete the enclosed procedure: Calibrating the Wash Buffet Level Sensor. Completing the enclosed procedure before using any of the affected lots of Part Number 8547197 will eliminate unnecessary workflow interruptions. Customers are to share the information with laboratory staff and retain this notification as part of their laboratory Quality System documentation. They are also instructed to complete and return the enclosed response form within 10 days so that the firm can be assured they were notified.

Device

  • 모델명 / 제조번호(시리얼번호)
    221208F (exp 6/9/07), 221209F(exp 6/9/07), 221210F (exp 6/10/07), 221211F (exp 6/10/07), 221212F (exp 6/12/07), 221213F (exp 6/12/07), 221214F (exp 6/13/07), 221215F (exp 6/13/07), 221216F (exp 6/14/07), 221217F (exp 6/14/07), 221218F (exp 6/15/07), 221219F (exp 6/15/07), 221220F (exp 6/16/07), 221221F (exp 6/16/07), 221222F (exp 6/17/07), 221223F (exp 6/17/07), 221224F (exp 6/19/07), 221225F (exp 6/19/07), 221226F (exp 6/20/07), 221227F (exp 6/20/07), 221228F (exp 6/21/07), 221229F (exp 6/21/07); All software versions
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide distribution --- including USA and country of Canada.
  • 제품 설명
    UniCel Dxl 800 Access Immunoassay Systems Wash Buffer, Part Number 8547197 Wash Buffer
  • Manufacturer

Manufacturer

  • 제조사 주소
    Beckman Coulter Inc, 200 S Kraemer Blvd, Brea CA 92822-6208