Antimitochondrial Antibody, Indirect Immunofluorescent, Antigen, Control - Product Code DBM
원인
The data reduction program written for a specific automated microtiter plate instrument (bio-tek elx800 reader) was found to be faulty. this program is used to calculate the cut off values for diastat anti-mitochondrial antibody kit. program should include a borderline specification to indicate repeat testing is needed.
조치
A notification letter and effectiveness check form was sent to consignees. This letter recommended that users manually verify results using the specifications outlined in the package insert until a new software version is installed.