FlexFlowTM venous cannula Catalog 200200 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Sorin Group USA, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    78331
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0076-2018
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Catheter, cannula and tubing, vascular, cardiopulmonary bypass - Product Code DWF
  • 원인
    A specific lot of the flexflowtm venous cannula, 200-200, is being recalled because it is sharper or pointier than expected.
  • 조치
    LivaNova sent an Urgent Medical Device Recall letter dated September 28 2017, to all affected consignees. The letter instructed customer to isolate products belonging to the lot involved. The level of the effectiveness check is Level A, where 100% of the consignees will be contacted. The customers were requested to send back any affected product and to complete the attached Customer Response Form by fax to 303-467-6502 or by email to USFSN@livanova.com. For questions customers were advised to contact Customer Service at 800-.650-2623.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot: 181185; UDI: (01)10803622125812(240)200-200(17)200703(10)181185
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US (nationwide) Internationally to France and Spain
  • 제품 설명
    FlexFlow Venous Cannula Model/Catalog Number 200-200 || The FlexFlow Venous Cannula is a single use sterile device. It is a size 23 French with an outside diameter of 0.30 inch (7.7 mm) and an effective working length of 11.2 inch (28.5 cm). The FlexFlow Venous Cannula is an open lumen polymer tube incorporating wire reinforcement in distal sections. The distal sections of the cannula are perforated with multiple openings to allow increased fluid flow. The clear proximal section is not reinforced to allow clamping the proximal end that terminates in a 3/8 inch (9.5 mm) barbed connector for standard cardiopulmonary bypass tubing. Each cannula is furnished with a mated obturator. Each obturator has a malleable wire. Each component is packaged inside a protective sheath in a single sterile, sealed pouch
  • Manufacturer

Manufacturer

  • 제조사 주소
    Sorin Group USA, Inc., 14401 W 65th Way, Arvada CO 80004-3503
  • 제조사 모회사 (2017)
  • Source
    USFDA