Immulite 2000 Rubella Quantitative IgG, Immulite 2500 Rubella IgG 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Siemens Healthcare Diagnostics Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    53675
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1033-2010
  • 사례 시작날짜
    2009-10-21
  • 사례 출판 날짜
    2010-03-10
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2010-04-22
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Rubella IgG - Product Code LFX
  • 원인
    The recall was initiated because siemens healthcare diagnostics has identified there is the potential for patient results to be biased low (-0.3 sample to cutoff [s/co] ratio units) due to incompatibility of the toxoplasma igm kits with this particular lot of igg/igm sample diluent. all available lots of the immulite 2000/2500 toxoplasma igm kits (l2ktm/l5ktm) are incompatible with diluent lot 150.
  • 조치
    The customer notification was initiated on 10/22/2009 with the firm forwarding an Urgent Field Safety notice with attached Field Correction Effectiveness Check form to the affected consignees informing them of the issue and the products affected. The consignees were informed that Siemens is initiating the Field Correction (FC) to prevent IMMULITE 2000/2500 Toxoplasma IgM from being run with Diluent lot 150. The consignees were instructed in the interim, until replacement kit lots of RUB IgG and CMV IgG are received, your laboratory can continue using Rubella IgG, CMV IgG, or Toxoplasma IgM as long as Diluent Lot 150 is not loaded on the analyzer when running Toxoplasma IgM. Diluent Lot 150 can be used with RUB IgG and CMV IgG. Once the replacements lots are received, any unopened kit lots listed in the above table MUST BE RETURNED to Siemens per instructions that will accompany a separate Fed Ex Call Tag that will explain the proper return of the affected product. The consignees were also instructed to complete the attached Field Correction Effectiveness Check form and fax to the firm's technical solutions center. Additional questions should be directed to the firm at 973-927-2828.

Device

Manufacturer

  • 제조사 주소
    Siemens Healthcare Diagnostics Inc., 5700 W 96th St, Los Angeles CA 90045-5544
  • 제조사 모회사 (2017)
  • Source
    USFDA