Antinuclear Antibody, Indirect Immunofluorescent, Antigen, Control - Product Code DHN
원인
A problem with software for the polytiter immunofluorescent titration system in that no flags or warnings appear when polytiter calibrator values are re-entered or changed (including values changed to unacceptable values) after the initial curve has been generated, if the 'refresh button' has not been pressed. the graph may appear valid; however erroneous results may occur.
조치
Customers were sent a Technical Bulletin describing the correct procedure to be used when entering data into the Poly/Titer tiration system, until the updated software is installed
OH, NJ, KY, TX, NC, FL, VA, WA, AL, CO, MO, AZ , CA , CT & NV
제품 설명
Programming software (all versions) for Polymedco PolyTiter Immunofluorescent Titration System and used with DiaSorin ANAFLUOR Indirect Fluorescent Antibody Test.