S1730 Calibration Solution 2, 200 mL, REF 944025. 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Radiometer America Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    35663
  • 사례 위험등급
    Class 3
  • 사례 연번
    Z-1460-06
  • 사례 시작날짜
    2005-10-31
  • 사례 출판 날짜
    2006-08-29
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-04-17
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Calibration solution - Product Code CHL
  • 원인
    Calibration solution for abl700 blood gas analyzers is labeled with an incorrect bar code. the bar code identified on the cal solution 2 product is actually the bar code for the rinse solution.
  • 조치
    Representatives of the recalling firm were informed of this problem and were asked to visit the firm''s customers and retrieve any of the suspect product found in inventory at each customer location for customer credit. The representatives were instructed to complete this action via a ''Management Cover e-mail issued on 10/31/2005. According to the Management Cover Memo, these actions were scheduled for completion by firm representatives on or before 12/15/2005.

Device

  • 모델명 / 제조번호(시리얼번호)
    Part #944-025, Lot QR-02, Expiration Date: September 2007.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    The recalled product was distributed to customers in following states nationwide: AL, AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV and WY
  • 제품 설명
    S1730 Calibration Solution 2, 200 mL, packed in single unit plastic bottles, REF 944-025, used with ABL700 Series Blood Gas Analyzers.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Radiometer America Inc, 810 Sharon Dr, Westlake OH 44145-1598
  • Source