Data reduction program written for specific sutomated microtiter plate insturment (bio-tek elx800 reader) is faulty in that the positive result limit is to be greater than 15 u/ml, rather than greater than or equal to 15 u/ml.
조치
Customers were sent notification letters and effectiveness check form. recommended users manually veify positve results (RSLT), using product insert specifications.
Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA assay kit.