Sopha Medical Vision DST / DSTXL Nuclear Cameras 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 General Electric Med Systems LLC 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    33832
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0105-06
  • 사례 출판 날짜
    2005-10-28
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2006-12-26
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System, Tomography, Computed, Emission - Product Code KPS
  • 원인
    Cracked welds on the main rotation shaft could potentially lead to the collapse of the gantry yokes that support the detectors if the system is continued to be used after the weld has completely fractured.
  • 조치
    GE Healthcare issued an Urgent Safety Notice, dated 07/22/05, to all affected sites. This letter describes the potential hazard, its waring signs and the additional corrective actions that will be taken by GE Healthcare.

Device

  • 모델명 / 제조번호(시리얼번호)
    Cameras affected by this safety notice are labeled with the SOPHA MEDICAL logo. STATIF DST and STATIF DST-XL model 100, serial numbers 001-264. The label is located on the inner side of the gantry cover. All affected cameras were manufactured prior to 1997 by Sopha Medial (SMV), which was subsequently acquired by GE Healthcare in 2000.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Within the US to include: AL, AZ, CA, DE, FL, GA, IA, IL, KS, MA, MD, MI, NC, NJ, NY, OH, PA, SC, TX, VA, WI and WV. OUS to include: Belgium, Brazil, Canada, China, Czech Republic, Egypt, France, Germany, Greece, Italy, Korea, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sinapore, Spain, Sweden and United Kingdom
  • 제품 설명
    Sopha Medical Vision DST / DST-XL Digital vatiable angle Nuclear Imaging System
  • Manufacturer

Manufacturer

  • 제조사 주소
    General Electric Med Systems LLC, 3000 N Grandview Blvd, Waukesha WI 53188-1615
  • Source
    USFDA