Toshiba America Medical Systems Inc Aplio Artida; 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Toshiba American Medical Systems Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    55466
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2179-2011
  • 사례 시작날짜
    2010-04-20
  • 사례 출판 날짜
    2011-05-10
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2011-05-25
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Diagnostic Ultrasound System - Product Code IYN
  • 원인
    There is a possibility of the system shutting down owing to the power supply unit causing the protection function to activate thus turning off the system prematurely.
  • 조치
    The recall was initiated on 4/20/10 with Toshiba America Medical Systems (TAMS) forwarding a Medical Device Correction with a Customer Reply Form (via US Postal Service for US Customers) to all customers who purchased the SSH-880CV Aplio Artida Diagnostic Ultrasound System. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to avoid this problem until corrective action is taken, in case this problem occurs; please contact their Toshiba Representative. If emergency use is necessary, use the following procedure which allows the system to be used temporarily: 1. Turn OFF the breaker at the back of the system and confirm that the error code near the breaker disappears. 2. Turn ON the breaker again to supply power to the system. Customers were instructed to complete and return the attached form and fax it to the toll free number at the top of the form. This form can also be sent via email to raffairs@tams.com. If customers had any questions they can call Paul Biggins at (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968.

Device

  • 모델명 / 제조번호(시리얼번호)
    W1F09X2272, W1F0972220, W1F0992260, W1F0992262, W1H1012311, W1F0972219, W1F09Y2281, W1F0972218, W1F09X2270, W1F09X2271, W1F09Y2280, W1F09X2268, W1F09X2273, W1F09Y2282, W1F09Z2296, W1F09Z2297, W1F0992261
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    USA, including Puerto Rico.
  • 제품 설명
    Toshiba America Medical Systems Inc. SSH-880CV Aplio Artida; Diagnostic Ultrasound System